| Instrument | Illuminance Spectrometer IM-1000 |
|---|
| JIS class | Conforms to the general AA class illuminance meter (JIS C 1609-1: 2006). |
|---|
| Spectral method | LVF(Linear Variable Filter) |
|---|
| Photodetector | Silicone photo diode array |
|---|
| Measurable wavelength range | 380 to 780 nm |
|---|
| Output wavelength resolution | 1 nm |
|---|
| Measurable illuminance range *¹ | 2 to 1,000,000 lx |
|---|
| Accuracy *¹ | Ev : ±2% ±1digit xy:±0.0020(50 lx or more) xy:±0.0035(10 to 50 lx) xy:±0.0050(5 to 10 lx) |
|---|
| Repeatability *¹・²・³ | Ev:0.5% + 1digit xy:0.0020(50 lx or more) xy:0.0035(5 to 50 lx) |
|---|
| Visible range relative spectral sensitivity characteristics (Difference from spectral relative luminous efficiency: f1’) | 2% or less |
|---|
| Systematic difference of angular incident light characteristics: f2 | 3% or less |
|---|
Temperature characteristics: fT | Within ±3% (-10 to 40℃ with 23℃ as reference) |
|---|
| Humidity characteristics: fH | Within ±3% (without dew condensation) |
|---|
| Measurement range mode | AUTO (AUTO FULL / AUTO FIRST / AUTO ADJUST) / MANUAL (MANUAL RANGE) |
|---|
| Display mode | XYZ / Ev / xy / u’v’ / Dominant wavelength λd / Excitation purity Pe / Correlated color temperature Tcp / Average color rendering property evaluation Ra / Special color rendering indexes Ri(i=1 - 15) / Spectral radiation illuminace graph / ⊿(XYZ) / ⊿(Ev、xy) / ⊿(Ev、u’v’) |
|---|
| Measurement time *⁴ | Approx. 0.2 seconds (When the measurement range is “MANUAL”, 100ms is fixed as the integral time and the “STR2” command is used) Approx. 0.5 to 50 seconds (Measurement range: AUTO) |
|---|
| Display | Liquid crystal display unit with 128×64 dots and back light ON/OFF function |
|---|
| Interface | RS-232C:Baud rate:9600 / 9200 / 38400bps, Parity:Odd number (ODD), Data length:7bit, Stop bit:1bit Wireless LAN:IEEE802.11b / 2.4GHz band(1-13ch) / 38400bps / Security(WEP / WPA / WPA2)(exclusively for Japan |
|---|
| Power supply | Nickel hydride AA battery: 4 pcs. (Standard accessory) / Exclusive AC adapter (optional accessory) *Battery life (Operable time): Approx. 7 hours |
|---|
| Operating condition | Temperature -10 to +40℃, Humidity 85%R.H. or less (without dew condensation) |
|---|
| External dimensions | Approx. 70 (W) × 250 (D) × 78 (H) mm (Without beam detector cap and power switch) |
|---|
| Weight | Approx. 640g (including the batteries) |
|---|
| Measurement reference surface | Edge of beam detector |
|---|